Cardiac Markers

[ Laboratory based EQA ]
Scheme Particulars Indicative Details
Accreditation body UKAS
ISO/IEC Accreditation status ISO/IEC 17043:2010
UKAS Accredited Proficiency Testing Provider No 8560
Objective of Scheme Provides EQA for laboratory based cardiac marker assays. Combined samples are issued for troponin, CKMB, myoglobin and NT-proBNP. An additional sample for troponin analysis only is issued for high sensitivity methods to cover very low clinically important levels around the assay detection limit.
Pre-, Analytical, Post- status Analytical
Number of distributions per year (nominal) 12
Specimens per distribution 4 (3 specimens for all analytes, +1 extra specimen to cover Troponin at low levels)
Number of Specimens per annum 48
Frequency of distribution Monthly
Scoring system ABC
Material distributed Lyophilised human serum
Programme Director(s) / Organiser(s) Dr Naomi Elkin
Criteria for participation All
Nominal sample volume 500 µL
Sample presentation TACs All analytes in the same tube
 
Analyte Approx Range Default Units Target Value B Score Limit (%) C Score Limit (%)
Cardiac Troponin I ~ ~ MLTM 15 15
Cardiac Troponin I (low levels) 3 – 20 ng/L MLTM 15 25
Cardiac Troponin T ~ ~ ALTM 15 15
Cardiac Troponin T (low levels) 3 – 20 ng/L ALTM 15 25
CKMB (Mass) ~ ~ ALTM 20 25
CKMB (Activity) ~ ~ ALTM 25 35
Myoglobin ~ ~ ALTM 20 25
NT-proBNP ~ ~ GLTM 15 20
 
Abbreviation Expansion
ALTM All Laboratory Trimmed Mean
GLTM Group Laboratory Trimmed Mean – which is the Method Principle Mean
MLTM Method Laboratory Trimmed Mean – usually a manufacturer system / analyser mean
XALTM An external value that may be from a Reference Method System
OGLTM A Method Principle Mean used as the target for all laboratories
OMLTM A specific Method Mean used as a target for all laboratories
MED Median